Proximal shaft: 1.73 F Distal shaft: 2.67 F
Crossing profile: 0.72 mm/0.028" ( 2.5 x 15 mm balloon)
Two Pipit SC balloon up to the size of 3.5 mm for all lengths, is 6F KBT compatible
Two Pipit SC balloon distal shaft size is 5.34 F, which is less than 5.4 F
Advanced multi-layered balloon material makes the balloon soft,with better shape memory
Upon deflation the balloon rewraps into a tri-fold shape
Pipit SC has a low rewrap profile of 1.84 mm
Enhanced crossability
Selective hydrophilic coating enhances the crossability
Effective and Precise dilatation
Minimizes slippage,prevents melon seeding
Enables precise dilatation
Balloon material | Multi-layered polyamide |
Balloon fold | Tri-fold |
Hydrophilic coating | Selective hydrophilic coating at distal shaft including balloon tip |
Balloon diameter | 2.0 mm - 4.5 mm |
Balloon length | 9.0 mm - 30 mm |
Tip entry profile | 0.016"/0.41mm |
Shaft diameter | Proximal 1.73 F (2.00 mm - 2.75 mm), 2.01 F (3.00 mm - 4.50 mm) Distal 2.67 F(2.00 mm-3.75 mm), 2.9 F ( 4.00 mm & 4.50 mm) |
Catheter length | 138 cm |
Rapid exchange length | 25 cm |
Markers on the proximal shaft | 85 cm and 95 cm |
Maximum recommended guidewire | 0.014"/ 0.36 mm |
Guide catheter compatibility | 5F compatible |
KBT compatibility | 6F ≤3.50 mm diameter and all lengths |
Pressure [atm] |
2.00 mm |
2.25 mm |
2.50 mm |
2.75 mm |
3.00 mm |
3.50 mm |
4.00 mm |
4.50 mm |
2 | 1.91 | 2.16 | 2.41 | 2.61 | 2.82 | 3.24 | 3.74 | 4.34 |
4 | 1.96 | 2.21 | 2.46 | 2.69 | 2.91 | 3.38 | 3.85 | 4.43 |
6 | 2.00 | 2.25 | 2.50 | 2.75 | 3.00 | 3.50 | 3.97 | 4.50 |
8 | 2.04 | 2.31 | 2.54 | 2.83 | 3.09 | 3.59 | 4.07 | 4.59 |
10 | 2.08 | 2.35 | 2.57 | 2.86 | 3.14 | 3.67 | 4.15 | 4.68 |
12 | 2.13 | 2.39 | 2.60 | 2.89 | 3.18 | 3.70 | 4.21 | 4.76 |
14 | 2.17 | 2.43 | 2.62 | 2.91 | 3.21 | 3.74 | 4.27 | 4.85 |
16 | 2.21 | 2.48 | 2.64 | 2.94 | 3.25 | 3.78 | 4.32 | 4.93 |
18 | 2.25 | 2.52 | 2.67 | 2.97 | 3.28 | 3.82 | 4.38 | 4.99 |
Diameter (mm) |
Length (mm) | |||||
9 | 12 | 15 | 20 | 25 | 30 | |
2.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
2.25 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
2.50 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
2.75 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
3.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
3.50 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
4.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
4.50 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
* upto 3.5 mm diameter for all lengths.
NP = Nominal Pressure, RBP = Rated Burst Pressure, KBT = Kissing Balloon Technique
Disclaimer: This product is intended for use by or under the directions of a physician. Prior to use, refer to the "Instructions for use" supplied with the product for indications, contraindications, side effects, suggested procedure warnings and precautions. As part of our continuous product development policy we reserve the right to change product specifications without prior notification.Pipit SC is a trademark of Sahajanand Medical Technologies Limited. Bench test results may not necessarily be indicative of clinical performance. Test performed by and Data on file at Sahajanand Medical Technologies Limited. Testing performed on Pipit SC (2.50 mm x 15 mm), Emerge SC (2.50 mm x 15 mm), Ryurei (2.50 mm x 15 mm), Trek (2.50 mm x 15 mm), Sapphire II Pro (2.50 mm x 15 mm). Information contained herein for distribution outside the USA, Japan & France only. Check the regulatory status of the device before distribution in areas where CE marking is not the regulation in force.
Data source: IFU/Marketing collaterals of respective brands. Tip and crossing profile of 2.50 mm balloon. Ryurei is a trademark of Terumo Corporation, Emerge SC is a trademark of Boston Scientific, Trek is a trademark of the Abbott group of companies, Sapphire II Pro is the trademark of Orbus Neich.